To test time measurement accuracy
To test time measurement accuracy
This test uses a precise frequency source to check the accuracy of time measurement
functions.
Specification
Delta-t accuracy
Repetitive (28 averages)
+[(0.005% x delta-t) + (0.1% of full scale)+ 100 ps]
Real Time * +/(0.005% x delta-t) + (0.2 x sample period)]
*
The specification applies to bandwidth limited signals (tr = 1.4 x sample period). The
sample period is defined as 1/(sample rate). The specification also applies to those automatic
measurements computing time intervals on identical slope edges (like pos-pos, neg-neg).
Equipment Required
Equipment
Critical Specifications
Recommended
Model/Part
Signal Generator
1-40 MHz,
HP 8656B Opt. 001
timebase accuracy - 0.25 ppm
Pulse generator
tr from 1.0 to 2.8 ns, 280 mVp-p,,externally
HP 8161A
triggerable
Cable
Type-N 24 inch
HP 11500B
Cable
BNC
HP 10503A
Adapter
Type N (f) to BNC (m)
HP 1250-0077
Repetitive Mode Procedure
This test checks time measurement in repetitive mode with averaging.
Set the HP 8656B signal generator to 40 MHz (25.0 ns period) and 800 mVrms.
Connect the output of the signal generator to the EXT INPUT of the HP 8161A pulse
generator.
Connect the output of the pulse generator to the channel | input.
Set up the pulse generator.
Input Mode = TRIG
Ext Input = 50 ohm, Trig Level = centered
Delay = 0s
Width = 12.5 ns
Leading Edge = 1.3 ns
Trailing Edge = 1.3 ns
High Level = 0.14 V
Low Level = -0.14 V
Press
Recall
,thenpress
cir
to set the oscilloscope to default conditions.
Press the channel 1 key (bottom row). Use the impedance soft key to select 50Q pe.
3-11