Repair; Assembly Replacement Procedures - Agilent Technologies N5532A User's And Service Manual

Sensor module
Hide thumbs Also See for N5532A:
Table of Contents

Advertisement

Repair

Tools required:
3/4-inch open end wrench
5/16-inch or 8 mm open end wrench (torque wrench preferred)
1/4-inch open end wrench
3/16-inch open end wrench
17 mm open end wrench
7 mm open end wrench
1.5 mm hex wrench
1/16-inch hex wrench
5/64-inch or 2 mm hex wrench
Torx - 10 driver
Needle-nose pliers
Diagonal cutters

Assembly Replacement Procedures

Remove and Replace the Instrument Covers
Use this procedure to remove and replace the instrument covers for all sensor
models.
1. Remove 4 screws and the bottom cover from the module assembly.
2. Locate the two aluminum deck blocks near the input connector. Notice that
there is a left and right side block, and how they fit. Remove the deck blocks
and separate the sensor module assembly from the plastic housing.
Figure 4-1
Deck Blocks
Chapter 4
Maintenance and Repair
Repair
23