Installation

Initial Inspection

Initial Inspection
Inspect the shipping container for damage. If the shipping container or packaging
material is damaged, it should be kept until the contents of the shipment have been
checked mechanically and electrically. If there is mechanical damage notify the
nearest Agilent Technologies office as shown in
damaged shipping materials (if any) for inspection by the carrier and an Agilent
Technologies representative.

Operating Precautions

NOTE
For optimal results, the sensor module must be placed on a supportive surface with
the serial number label facing up. This orientation provides repeatable
measurements by eliminating strain on the input connector.
The input connector should be torqued to 12 in-lb (135 Ncm) for the type-N
connector, 8 in-lb (90 Ncm) for the APC 3.5 mm connector, or 8 in-lb (90 Ncm)
for the 2.4 mm connector to ensure a repeatable connection in order to meet the
published specifications. Exceeding these torque settings may damage the
connector.
Connect the sensor module by turning only the hex nut portion of the connector.
Damage can occur if torque is applied to the sensor module body.
The connector plastic insulator bead deteriorates when contacted by acetone,
trichloroethylene, carbon tetrachloride, benzene, etc.
BEFORE CONNECTING THE SENSOR MODULE TO OTHER
WARNING
INSTRUMENTS, ensure that all instruments are connected to the protective
(earth) ground. Any interruption of the protective earth grounding will cause
a potential shock hazard that could result in personal injury and cause
damage to the sensor module.