SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
Connecting/Disconnecting the Cylinder
to/from the Compressor
Connecting the Cylinder to the Compressor
NEVER use tools of any kind to connect/disconnect the
cylinder and the compressor. Otherwise, severe injury
and/or damage may occur.
DO NOT drop oxygen cylinders. Use two hands when
handling/transporting oxygen cylinders. Otherwise, injury
or damage may occur.
DO NOT use a liquid leak detector to test for leaks.
NOTE: For this procedure, refer to FIGURE 4.2 and FIGURE 4.3.
1. Examine the cylinder gauge. If the cylinder pressure is less 
than 10342 kpa (1500 psig), proceed to STEP 2 (FIGURE 
4.3).
NOTE: If the cylinder pressure is greater than 10342 kpa (1500 
psig), DO NOT attempt to top off this cylinder. It may not fill.
2. Set the regulator flow dial on the cylinder to Off (FIGURE 
4.3).
3. Remove the cylinder and connector fillport covers.
DO NOT connect the outlet (cannula) connection to the
HomeFill II unit, otherwise the conserving cylinder will
not work properly.
4. Momentarily push DOWN on the outer ring (sleeve) of the 
connector fillport to reset the connector as shown in 
FIGURE 4.2.
HomeFill® II
WARNING
CAUTION
16
Part No 1141477