Franklin™ Real-Time PCR Thermocycler (EUA) For IVD Use ONLY. For Emergency Use Authorization ONLY. This product has not been FDA cleared or approved, but has been authorized by FDA under an EUA for use by authorized laboratories that are certified under the Clinical Laboratory Improvement Amendments of 1988 (CLIA), 42 U.S.C.
Franklin™ Thermocycler The Biomeme Franklin™ transforms your smartphone into a thermocycler for real-time PCR analysis with results in under an hour. Our mobile thermocycler enables multiplex real-time detection of up to 27 targets from 1 sample or test 9 samples for up to 3 targets each. Just under 3 pounds, hand-held, and battery- operated for maximum portability enabling a full day’s work out in the field on a...
LED Status Indicators Vertical LED on the front of your thermocycler. The Franklin™ thermocycler has 5 LEDs on the front of the unit. The LEDs are used to convey various states of the thermocycler as outlined in the table below.
Once enabled, tap Connect via BLE when prompted in the smartphone app. If working with multiple Franklin™ thermocyclers, select the appropriate unit and tap Confirm. The LED on the front of your thermocycler will flash white indicating it’s connected.
Page 11
“My run has completed, and I’ve pressed a bunch of buttons. I’m feeling flustered, confused, and frustrated; what should I do?” 1. Turn off the thermocycler and close the Biomeme Go Dx app. 2. Follow the steps described in Scenario below.
Page 12
Bluetooth back on. 3. Within the Biomeme Go Dx app, you should see a pop-up screen that reads: “Connect via BLE”. If so, press Scan in the Biomeme Go Dx app. If not, power OFF the thermocycler, CLOSE the app, and to Scenario below.
Page 13
C: Thermocycler OFF / App OPEN 1. Press the Power button on top of your Franklin™ thermocycler to turn it back on. If your thermocycler doesn’t turn back on, make sure it’s connected to a power source as your battery may have died.
5. Relaunch the Biomeme Go Dx app by selecting the icon on your smartphone’s home screen. 6. From the app’s home screen, tap Incomplete Runs. 7. Select your test from the list of incomplete runs. 8. Press the Scan button in the app and select your thermocycler.
Go-Strip wells. Tap the bottom of the sealed Go-Strip gently but firmly against a solid surface before removing the foil seal and adding your sample. 1. Open the contents of a Biomeme SARS-CoV-2 Go-Strips pouch. Do not immediately discard the Go-Strips pouch as you’ll need to scan the QR code in a later step.
3. Close the thermocycler lid securely and navigate to the Biomeme Go Dx mobile application on your smartphone to begin your testing protocol. For further instructional information, please contact support@biomeme.com.
Page 18
Samples & Controls Note: Transport your Franklin™ thermocycler in its carrying case. Additionally, moving your thermocycler while thermocycling could result in errors. We highly recommend not moving or opening the device while thermocycling to avoid losing your PCR run. After your run has completed, be careful when removing your Go-Strips and void filling caps to avoid liquid splatter and PCR amplicon contamination.
The outside of the Franklin™ thermocycler can be cleaned using a 70% ethanol solution which must be sprayed on a cloth rather than directly on the Franklin™ thermocycler. Lysol wipes or Micro-Chem Plus wet paper towels are acceptable as well.
If your test fails, the smartphone will notify you of the error returned from the thermocycler. Your last run will be saved in the Biomeme mobile app up to the point of failure, but the data will not be processed resulting in no Cq values, baseline, or graph of smooth data.
If your run fails to start, the app will return to the home screen and have you restart the setup of your test. Restart the thermocycler then reconnect the smartphone. If starting still fails after many retries, please contact support@biomeme.com. The USB failed to send the protocol...
Page 26
564(b)(1) of the Act, 21 U.S.C. § 360bbb-3(b)(1), unless the declaration is terminated or authorization is revoked sooner. Biomeme products may not be transferred to third parties, resold, modified for resale or used to manufacture commercial products or to provide a service to third parties without written approval of Biomeme, Inc.