Technical Data

Flow Rate Settings
Oxygen Concentration
Oxygen Concentration Indicator
Maximum System Pressure
Oxygen Outlet Pressure
Electrical Power:
AC Operation
DC Operation
Power Cartridge
AC Power Supply
DC Power Supply
Operating Temperature
Operating Humidity
Nominal Sound Level
Dimensions (H x W x D)
Weight
Continuous Flow Indication
Audible Alarm Indicators
Back-Up Alarm Power
Filters
Device Classification
Transport/Storage Requirements
Back pressure Effect on Flow Indicator
Altitude
38 | ECLIPSE
TM
Model 1000
User Manual
*
1 to 6 setting in Pulse Flow Mode
0.5 to 3.0 LPM in Continuous Flow Mode
90% +/– 3% for all flow settings
Green Light = Normal Operation
Yellow Light = Warning or Caution, less than 85% +/– 3%
Red Light Flashing = Abnormal Operation, less than 70% +/– 5%
15 psig (103.5 kPa)
5.0 psig (34.5 kPa) nominal
100 – 240 VAC, 50/60 Hz
12 V
195 W-Hrs
200 Watts Max
150 Watts Max
50° F to 104° F (10° C to 40° C)
10% - 95% at 82.4° F (28° C) dew point
48 dBA at 3.0 LPM in Continuous Flow Mode
40 dBA at 3.0 LPM in Pulse Flow Mode
19.3 inches x 12.3 inches x 7.1 inches
(49.0 cm x 31.2 cm x 18.0 cm)
17.9 pounds (8.1 kg) with Power Cartridge
14.5 pounds (6.6 kg) without Power Cartridge
Expressed in liters per minute (LPM)
Loss of power/Hot Power Cartridge
Low Battery/Warm Power Cartridge
Low Oxygen Output
O
flow outside normal limits
2
No inspiration detected in Pulse Mode
Unit Malfunction
9V Internal Battery
Air Inlet, HEPA
IEC Class I, Type B Applied Part, IPX0
Temperature: -4° F (-20°C) to +140° F (+60° C)
Humidity: up to 95% Non-Condensing
None
0 –13,123 feet (0 – 4,000 meters)