Alarm Conditions and Troubleshooting

The table shows possible alarm conditions and suggests appropriate responses. If any other alarms occur,
call your authorized SeQual dealer/provider.
Condition
"Warming Up. Please Wait. "
"2.0" (or flow setting)
"Low or Warm Power
Cartridge"
"O2 Concentration < 85%" Off
"O
Concentration < 70%"
2
"Flow Rate Error/
Blocked Flow"
"No Inspiration Detected"
(in Pulse Flow Mode only)
"Power Cartridge Shutdown
or hot Power Cartridge" (with
no external power present)
"Loss of External Power"
(without charged Power
Cartridge installed)
"System Malfunction"
AVAILABILITY OF AN ALTERNATE SOURCE OF OXYGEN IS
WARNING
RECOMMENDED IN CASE OF POWER OUTAGE OR MECHANICAL
FAILURE. CONSULT YOUR PHYSICIAN FOR THE TYPE OF
BACK-UP SYSTEM REQUIRED.
30 | ECLIPSE
Model 1000
User Manual
TM
Green
Yellow
Red
Indicator
Indicator
Indicator
On
On
On
On
Off
Off
On
Blinking
Off
Blinking
Off
Off
Off
Blinking
Off
Blinking
Off
On
On
Off
Off
Off
Blinking
Off
Off
Blinking
Off
Off
On
Audible
Alarm
What to do
None
Wait. The system typically takes
3–5 minutes to reach specified
performance.
None
Nothing. The system is operating
properly at the specified flow rate.
Beeps
Plug into external power or replace
with a charged Power Cartridge.
None
Continue Use. Check or clean air
intake filter. If condition persists,
contact dealer/provider.
Beeps
Continue Use. Check or clean air
intake filter. If condition persists,
contact dealer/provider.
Beeps
Check tubing and/or humidifier
for obstruction. Check/clean
air intake filter. If condition
persists past 10 minutes,
contact dealer/provider.
Beeps
Make sure tubing does not
exceed 7 ft (2.1m) when using
Pulse Flow Mode. Audible alarm
will beep after 15 seconds if no
inspiration is detected. If no
inspiration detected 15 seconds
after that, the ECLIPSE will
automatically switch to
Continuous Flow Mode.
On
Seek external power source.
If Power Cartridge malfunction
persists, contact dealer/ provider.
On
Seek an alternative oxygen
supply until external power is
restored or install a charged
Power Cartridge.
On
Contact dealer/provider. Use
back up oxygen supply.
DO NOT IGNORE ALARMS.