Recommended Operating Environments (continued)

The Proper Location

Select a location that avoids fumes and pollutants and allows the oxygen system to
draw in room air through the air inlet filter on the back of the cabinet and exhaust air
out the vent at the bottom of the case without being restricted.
Locate the unit such that the alarms can be heard.
Route oxygen tubing such that it does not kink or is subject to occlusion.
DO NOT cover unit with a blanket, towel, quilt, or other covering, as the unit may
overheat.
DO NOT locate near flammable materials or cleaning agents or in the direct path of
any heat source, such as a heat register or car heater.
Keep the oxygen system at least five (5) feet (1.5 m) away from hot, sparking objects
or open sources of flame.