Model 1000 User Manual

Advertisement

Table of Contents
loading

Summary of Contents for Eclipse 1000

  • Page 1 Model 1000 User Manual...
  • Page 3: Table Of Contents

    Indications For Use Contraindications Safety Guidelines Oxygen Therapy Safety Operating Safety Power Cartridge Safety Introduction To Your ECLIPSE Oxygen System Understanding the Controls and Front Panel Description of the ECLIPSE Control Panel Features User Controls Front Panel Icons System Status Indicators...
  • Page 4: Important Safety Instructions

    Technologies Inc. (SeQual) recommends that this User Manual be given to each patient receiving an ECLIPSE Oxygen System. This User Manual describes all the necessary information in order for a patient or caregiver to operate the ECLIPSE. Important safety information for hazards that might cause WARNING serious injury.
  • Page 5 The ECLIPSE enclosure does not “Drip Proof Equipment-IPX0”: provide protection against the harmful effects of liquid ingress. (IPX0, per IEC 60529) The ECLIPSE AC and DC Power “Drip Proof Equipment-IPX1”: Supplies provide protection against the harmful effects of liquid ingress. (IPX1, per IEC 60529) “IEC Icon for DC Power (Input)”...
  • Page 6 Shows the amount of charge remaining in “Power Cartridge Status”: the Power Cartridge. The symbol on the Power Cartridge “Power Cartridge Symbol”: pack. The port in which the oxygen flows out and where “Oxygen Output”: the tubing is attached. 4 | ECLIPSE Model 1000 User Manual...
  • Page 7: Indications For Use

    Model 1000 User Manual Indications for Use THE ECLIPSE OXYGEN SYSTEM IS INDICATED FOR THE ADMINISTRATION OF SUPPLEMENTAL OXYGEN. THE DEVICE IS NOT INTENDED FOR LIFE SUPPORT NOR DOES IT PROVIDE ANY PATIENT MONITORING CAPABILITIES. YOUR PHYSICIAN HAS PRESCRIBED A SPECIFIC OXYGEN FLOW RATE SETTING TO MEET YOUR INDIVIDUAL NEEDS.
  • Page 8: Safety Guidelines

    Keep these substances away from the oxygen system, tubing and connections and any other oxygen source. DO NOT use any petroleum based or other lubricants. 6 | ECLIPSE Model 1000 User Manual...
  • Page 9: Operating Safety

    DO NOT drop or insert any object into any opening. DO NOT block the air inlet or the exhaust vent of the ECLIPSE when it is on a soft surface, such as a bed or couch or car seat.
  • Page 10 Locate oxygen tubing and Power Supply cords to prevent tripping CAUTION hazards. DO NOT place ECLIPSE in a small, enclosed space, such as a CAUTION bathroom or closet, with cannula leading out of the enclosed space. 8 | ECLIPSE...
  • Page 11: Power Cartridge Safety

    Model 1000 User Manual Power Cartridge Safety This Power Cartridge may explode and cause potential injury WARNING if exposed to or disposed of in a fire. short-circuit the Power Cartridge’s metal contacts with WARNING DO NOT metallic objects, such as keys or coins. It may cause sparks or excessive heat.
  • Page 12: Introduction To Your Eclipse Oxygen System

    SUITS THE PATIENT’S LIFESTYLE. THE ECLIPSE MAY BE USED IN HOME, INSTITUTION, AND MOTOR VEHICLES. USE OF THE ECLIPSE IN HOSPITALS SHOULD INCLUDE THE USE OF A “HOSPITAL GRADE” POWER CORD. YOUR PHYSICIAN HAS PRESCRIBED A SPECIFIC OXYGEN FLOW RATE SETTING TO MEET YOUR INDIVIDUAL NEEDS.
  • Page 13: Understanding The Controls And Front Panel

    Model 1000 User Manual Understanding the Controls and Front Panel Description of the ECLIPSE The key features of the ECLIPSE Oxygen System and control panel are important items to become familiar with. Handle Control Panel Oxygen Outlet Port Service Port (Not for Patient Use) Simplified User...
  • Page 14: Control Panel Features

    Understanding the Controls and Front Panel (continued) Control Panel Features The ECLIPSE displays important operating information using the icons, numbers and visual indicators on the control panel. This section will help you understand these features. No Smoking Icon ON/OFF Button...
  • Page 15: User Controls

    Model 1000 User Manual User Controls This section describes the control buttons and function for your ECLIPSE Oxygen System. F R O N T PA N E L I C O N S This button turns the system ON or ON/OFF Button and Indicator: OFF.
  • Page 16 If the Power Cartridge is not installed, or if it is improperly installed, the Status Gauge will not light. When the ECLIPSE system is External Power Present Indicator: properly plugged in using the AC or DC Power Supply, this indicator will appear on the Control Panel.
  • Page 17: Recommended Operating Environments

    Operating Time Up to 24 hours per day when connected to external power. When moving the ECLIPSE from an extreme environment, allow time for your ECLIPSE to acclimate to the recommend operating environment. Operating your ECLIPSE outside the recommended operating...
  • Page 18: The Proper Location

    DO NOT locate near flammable materials or cleaning agents or in the direct path of any heat source, such as a heat register or car heater. Keep the oxygen system at least five (5) feet (1.5 m) away from hot, sparking objects or open sources of flame. 16 | ECLIPSE Model 1000 User Manual...
  • Page 19: Operating Instructions

    If you have any further questions or problems call your SeQual dealer/provider or authorized service representative. DO NOT attempt to operate the ECLIPSE System without first reading Important! the “Safety Guidelines” section of this manual. Follow all operating instructions.
  • Page 20: Powering Your Eclipse System

    (1) AC POWER SUPPLY (2) DC POWER SUPPLY The ECLIPSE charges the optional Power Cartridge when external power (AC or DC) is present and when the Power Cartridge temperature is below a safe charging temperature. When external power is disconnected, the system will automatically switch over to the Power Cartridge, if installed.
  • Page 21 Insert the DC Start your vehicle. cord in the proper receptacle on the side of the ECLIPSE. Next, insert the DC Power Supply plug into the DC accessory outlet in your vehicle. When the system is properly plugged in and receiving power from the DC source, a green indicator...
  • Page 22 Operating Instructions (continued) Secure the ECLIPSE and the DC Power Supply in your vehicle and WARNING make sure the air inlet and exhaust airflow are not blocked. This may impact performance. leave the ECLIPSE or DC Power Supply plugged into the...
  • Page 23 A variety of factors, such as flow rate and Pulse or Continuous Flow Mode operation determine the duration of operating time. The following table provides estimates of amount of time that the ECLIPSE can operate on a fully charged Power Cartridge given certain flow rates and operating conditions.
  • Page 24 DO NOT disassemble, incinerate, or heat above 140° F (60° C). Return your Power Cartridge to your dealer/provider for proper CAUTION disposal. Keep your Power Cartridge away from children. CAUTION 22 | ECLIPSE Model 1000 User Manual...
  • Page 25 I N I T I A L P OW E R CA R T R I D G E O P E R AT I O N : The new Power Cartridge packaged with your ECLIPSE Oxygen System is not fully charged. Before using your ECLIPSE Oxygen System for the first time, you must fully discharge the Power Cartridge.
  • Page 26 Use of another Power Cartridge or Battery may damage the unit, present a risk of fire or explosion and will void the warranty. In the event of a power interruption, the ECLIPSE Oxygen System will automatically switch to the Power Cartridge operation. When power is restored, the Power Cartridge will automatically start recharging.
  • Page 27: Operating Your Eclipse System

    Model 1000 User Manual Operating Your ECLIPSE STEP 1: LOCATING YOUR ECLIPSE Place the ECLIPSE in a well-ventilated place so that the air inlet and exhaust have clear access. Position the ECLIPSE so that any alarm can be heard. Check to see if the air inlet filter is in place before using. If the air inlet filter needs to be cleaned, wash with soapy water and allow drying before using.
  • Page 28 Press and hold the Button for one (1) second to turn “ON/OFF” your ECLIPSE. A brief self-test will happen. Verify that all the indicators are lit and a 3-second buzzer sounds, indicating the system is in working order. Warm-up Period •...
  • Page 29 Model 1000 User Manual STEP 3: CONNECT THE TUBING TO THE OXYGEN OUTLET AND THE CANNULA PER THE INSTRUCTIONS FROM YOUR DEALER/PROVIDER. Replace the cannula and tubing on a regular basis. Consult your dealer/provider for cannula and tubing cleaning and replacement information.
  • Page 30 When operating in Pulse Flow Mode, to ensure adequate breath WARNING detection, the length of the tubing/cannula must not exceed 7 feet (2.1m). As with all conserving devices, the ECLIPSE may not be able to detect WARNING some respiratory efforts. STEP 5: ADJUST FLOW RATE TO PRESCRIBED LEVEL Using the , adjust the flow rate to...
  • Page 31 The Cart is designed for use on pedestrian surfaces. Align the locator holes on the bottom of the unit to the tabs on the cart. Align the screw to the insert on the back of the ECLIPSE and hand-tighten the knob on the cart.
  • Page 32: Alarm Conditions And Troubleshooting

    7 ft (2.1m) when using Pulse Flow Mode. Audible alarm will beep after 15 seconds if no inspiration is detected. If no inspiration detected 15 seconds after that, the ECLIPSE will automatically switch to Continuous Flow Mode. “Power Cartridge Shutdown Blinking Seek external power source.
  • Page 33 Model 1000 User Manual Symptom Probable Cause Remedy ECLIPSE does not power on No Power Cartridge installed Install charged Power Cartridge when ON/OFF button is pressed or plug into external power Power Cartridge is discharged Plug into external power Malfunction...
  • Page 34: Accessories

    Use of humidifiers not specified for this oxygen system may impair performance and void the warranty. DO NOT use a humidifier in Pulse Flow Mode. The ECLIPSE will not detect an inspiration and will default to Continuous Flow Mode.
  • Page 35 ECLIPSE is capable of providing you with battery operation depending on your prescribed flow rate. Your Power Cartridge will be recharged when the ECLIPSE is connected to the AC Power Supply or the DC Power Supply, if adequate power is available. The Power Cartridge will take longer to recharge when connected to the DC Power Supply and will depend on the flow setting and the amperage switch setting.
  • Page 36 WARNING DO NOT shock or serious injury may result. Your ECLIPSE Oxygen System can be easily transported while attached to the Cart: Cart. The cart has wheels and a telescoping handle for easy mobility. DO NOT drip, spill or expose the ECLIPSE to water during portable WARNING use.
  • Page 37: Cleaning, Care And Routine Maintenance

    Your dealer/provider will perform inspections and maintenance of the compressor, filters, alarms, 9V battery, Power Cartridge and other internal parts as needed. disassemble the ECLIPSE or attempt to perform any routine WARNING DO NOT maintenance other than what is described in the Patient Cleaning and Care section below.
  • Page 38 Turn OFF the Cabinet, AC Power Supply, DC Power Supply and Power Cord: ECLIPSE and disconnect from AC or DC power before any cleaning or disinfection activity. Use a damp (not soaking wet) cloth or sponge and a mild detergent solution to clean the cabinet and power supplies.
  • Page 39 Model 1000 User Manual Power Cartridge Cleaning, Care and Maintenance The Power Cartridge in the ECLIPSE Oxygen System requires special care to assure a longer life and higher performance. The SeQual Power Cartridge is the only approved Power Cartridge to be used.
  • Page 40: Technical Data

    IEC Class I, Type B Applied Part, IPX0 Transport/Storage Requirements Temperature: -4° F (-20°C) to +140° F (+60° C) Humidity: up to 95% Non-Condensing Back pressure Effect on Flow Indicator None Altitude 0 –13,123 feet (0 – 4,000 meters) 38 | ECLIPSE Model 1000 User Manual...
  • Page 41 6 to 25 breaths per minute average, sustainable Minimum time between breaths 1.25 seconds (maximum 3 consecutive breaths) Response to missing breaths ECLIPSE switches to Continuous Flow Mode after 30 seconds without breath detection. Specifications are subject to change without notice. | 39...
  • Page 42: Summary Of All Warnings And Cautions

    Summary of All Warnings and Cautions FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE WARNING ORDER OF A PHYSICIAN. THE ECLIPSE IS NOT INTENDED FOR LIFE SUPPORTING OR LIFE WARNING SUSTAINING APPLICATIONS, NOR DOES IT PROVIDE ANY PATIENT MONITORING CAPABILITIES.
  • Page 43 DO NOT operate device or accessories on wet surfaces or in standing WARNING water and DO NOT submerse or expose to water. The ECLIPSE enclosure does not provide protection against the harmful effects of liquid ingress. Electrical shock or damage to the unit may result.
  • Page 44 Cartridge removed while the unit is connected to an external power source. This may result in electrical shock. Secure the ECLIPSE and the DC Power Supply in your vehicle and WARNING make sure the air inlet and exhaust airflow are not blocked. This may impact performance.
  • Page 45 WARNING detection, the length of the tubing/cannula must not exceed 7 feet (2.1m). As with all conserving devices, the ECLIPSE may not be able to detect WARNING some respiratory efforts. DO NOT change from prescribed flow rate setting without consulting WARNING with a qualified clinician.
  • Page 46 Exposing the Power Cartridge to water or other liquids may cause WARNING personal injury. DO NOT drip, spill or expose the ECLIPSE to water during portable WARNING use. The ECLIPSE enclosure does not protect against the harmful effects of liquid ingress. Electrical shock or damage to the unit may result.
  • Page 47 Locate oxygen tubing and Power Supply cords to prevent tripping CAUTION hazards. DO NOT place ECLIPSE in a small enclosed space, such as a CAUTION bathroom or closet, with cannula leading out of the enclosed space. DO NOT expose the Power Cartridge to temperatures above 140° F CAUTION (60°...
  • Page 48 The AC power cord used with the ECLIPSE Oxygen System must CAUTION meet the requirements of the country where the product is used. Use with grounded, 3-pin receptacle. DO NOT remove ground pin CAUTION from cord. Use only with SeQual supplied cords.
  • Page 49 CAUTION Store the Power Cartridge in a cool, dry place when not in use. CAUTION DO NOT leave your ECLIPSE or your Power Cartridge in your vehicle CAUTION or in a trunk during a hot day. The ECLIPSE system can only work with a SeQual Power Cartridge.
  • Page 50: Electromagnetic Compatibility

    Consult the manufacturer or service technician for help. Guidance and Manufacturer’s declaration - electromagnetic emissions The SeQual Technologies ECLIPSE Oxygen System is intended for use in the electromagnetic environment specified below. The customer or the user of the SeQual Technologies ECLIPSE Oxygen System should assure that it is used in such an environment.
  • Page 51: Disposal Of Equipment And Accessories

    Model 1000 User Manual Disposal of Equipment and Accessories Follow local governing ordinances and recycling plans regarding disposal of device components. DO NOT dispose of the Power Cartridge. Power Cartridges should be returned to your dealer/provider. MANUFACTURED BY: EUROPEAN UNION AUTHORIZED AGENT: SeQual Technologies Inc.
  • Page 52 ECLIPSE Model 1000 User Manual PN 6112 03/06...