Verifying The Scxi-1502 Installation In Software; Installing Scxi Chassis And Modules In Software; Troubleshooting The Self-Test Verification; Troubleshooting In Ni-Daqmx - National Instruments SCXI-1502 User Manual

Table of Contents

Advertisement

Verifying the SCXI-1502 Installation in Software

Installing SCXI Chassis and Modules in Software

Troubleshooting the Self-Test Verification

Troubleshooting in NI-DAQmx

© National Instruments Corporation
Refer to the SCXI Quick Start Guide for information on verifying the SCXI
installation.
Refer to the SCXI Quick Start Guide for information on installing chassis
and modules using NI-DAQmx in software.
If the self-test verification did not verify the chassis configuration,
complete the steps in this section to troubleshoot the SCXI configuration.
If you get a Verify SCXI Chassis message box showing the SCXI
chassis model number, Chassis ID: x, and one or more messages
stating Slot Number: x Configuration has module: SCXI-XXXX
or 1502, hardware in chassis is: Empty, take the following
troubleshooting actions:
Make sure the SCXI chassis is powered on.
Make sure all SCXI modules are properly installed in the chassis.
Refer to the SCXI Quick Start Guide for proper installation
instructions.
Make sure the cable between the SCXI chassis and E/M Series
DAQ device is properly connected.
Inspect the cable connectors for bent pins.
Make sure you are using the correct NI cable assembly.
Test the E/M Series DAQ device to verify it is working properly.
Refer to the E/M Series DAQ device help file for more
information.
1-5
Chapter 1
About the SCXI-1502
SCXI-1502 User Manual

Advertisement

Table of Contents
loading

Table of Contents