Maintenance
Maintenance
2
[374|376|378|703|704|634|635|644|645|656|657|664|665|674|675
|676|677|684|685|264|274|284|285|954|957|959|963]
The following section provides an overview of all maintenance jobs which can
be carried out.
To call up the Service Menu, start the scale while pressing another key at the
same time. This causes the content of the calibration counter (i.e. the number of
adjustment operations carried out so far) to be displayed for 24 seconds
(flashing). While the calibration count is displayed, a key must be kept pressed
(for more than 1.5 seconds); the scale then automatically displays the software
identification number and the check digit.
Keypress (any key) and START key simultaneously
-
Calibration counter [see 2.1.2]
-
Long keypress (any key)
Automatic display for 6 seconds
-
XX.YY (software identification)
Then automatic display for 6 seconds
-
XXXX (check digit)
Then automatic display
-
CAL [see 2.1.5]
Long keypress (any key)
o DEC
Long keypress (any key)
o INC
Short keypress (any key)
-
GAL [see1.3]
Long keypress (any key)
o XXXXX (gravity factor)
Short keypress (any key)
-
INFO
14.11.12/RLEU/ARI/MRE
8/40
seca
30-34-00-812c