For example, if the Corrected Insertion Loss is −10.0 dBm, and the
measuring receiver reading is −30 dBm, then −30 dBm − (−10.0)
dBm = −20 dBm.
32.Adjust A9R19 CAL OUT ADJ accordingly if the CAL OUT amplitude
is not −20 dBm ±0.4 dB as calculated in
Chapter 2
13. Third Converter and Second IF Bandpass
CAL OUT Power____________________dBm
Making Adjustments
step
26.
91