Agilent Technologies 8590 Series Service Manual page 130

Analyzers assembly-level repair
Table of Contents

Advertisement

Making Adjustments
19. Entering External ALC Correction Constants for Option 010 or 011
4. Press the following analyzer keys.
CAL, More 1 of 4
EXECUTE TITLE CAL
,
SERVICE DIAG
Verify that the XSLOP and XOFST values are the same as you
entered in
step
3.
5. Perform the "Checking the External ALC for Option 010 or 011"
procedure.
130
,
,
More 2 of 4
SERVICE CAL
,
,
More 1 of 4
More 2 of 4
,
DISPLAY CAL DATA
NEXT PAGE
,
Chapter 2