Agilent Technologies 8590 Series Service Manual page 340

Analyzers assembly-level repair
Table of Contents

Advertisement

Replacing Major Assemblies
A17 Memory Card Reader Assembly
Replacement
1. Place the analyzer on the worksurface with the bottom side facing
2. Place the A17 Memory Card Reader assembly on the memory-card
3. Replace the four screws (4) that secure the A17 assembly to the
The two alignment tabs on the memory-card base beneath the A17
CAUTION
Memory Card Reader assembly can be easily damaged. When replacing
the A17 Memory Card Reader assembly, carefully mate the two
alignment tabs with the memory-card bezel on the front-frame
assembly.
4. Replace the A17 assembly (3), gently sliding the assembly along the
5. Replace the two screws (2) that secure the A17 assembly to the
6. Replace the jumper (5) between the A17 Memory Card Reader and
7. Connect the W14 ribbon cable (1) to the A17 assembly.
8. Replace the instrument cover assembly. Refer to the
340
up.
base.
memory-card base.
chassis toward the memory-card reader bezel on the front-frame
assembly.
chassis.