Equipment Required
Synthesized sweeper
Measuring receiver (used as a power meter)
Power sensor, 50 MHz to 26.5 GHz
Power sensor, 1 MHz to 2.9 GHz
Power splitter
Adapter, Type N (m) to APC 3.5 (m)
Adapter, Type N (f) to APC 3.5 (m)
Adapter, APC 3.5 (f) to APC 3.5 (f)
Cable, APC 3.5, 91 cm (36 in.)
Procedure
Measuring Uncorrected Flatness
1. Zero and calibrate the measuring receiver and 1 MHz to 2.9 GHz
power sensor in log mode as described in the measuring receiver
operation manual.
2. Connect the equipment as shown in
Figure 2-30
Frequency Response Setup
Chapter 2
16. Frequency Response of the 8592L/94L, 8593E/94E/95E/96E, and 8594Q
Making Adjustments
Figure
2-30.
107