Troubleshooting And Maintenance; Troubleshooting; Maintenance - Campbell 108 Instruction Manual

Hide thumbs Also See for 108:
Table of Contents

Advertisement

Model 108 Temperature Probe
9.

Troubleshooting and Maintenance

NOTE
9.1

Troubleshooting

9.2

Maintenance

10
snip examples increase settling time by 20000 µs by placing 20000 as the
argument in the SettlingTime parameter:
CR6-series datalogger example:
Therm108(T108_C,1,U1,U10,20000,60,1.0,0.0)
CR800-series and CR3000 dataloggers example:
Therm108(T108_C,1,1,1,20000,_60Hz,1.0,0.0)
All factory repairs and recalibrations require a returned material
authorization (RMA) and completion of the "Declaration of
Hazardous Material and Decontamination" form. Refer to the
page at the beginning of this manual for more
Assistance
information.
Symptom: Temperature is reported as NAN, –INF, or incorrect temperature.
Verify wire leads are connected to the terminals specified in the
Therm108() instruction: red to single-ended analog input (SE or U), black
to switched excitation (VX/EX or U), and purple to ground (⏚).
Symptom: Incorrect temperature is reported.
Verify the Mult and Offset arguments in
desired units (Section 7.2, Datalogger Programming
for signs of damage and possible moisture intrusion.
Symptom: Unstable temperature is reported.
Probably a result of electromagnetic interference. Try using the 50 Hz or
60 Hz options for the Fnotch parameter, and/or increasing the settling time
as described in Section 8.3, Electrically Noisy Environments
Section 8.4, Long Cable Lengths
datalogger ground, and the datalogger is properly grounded.
The 108 probe requires minimal maintenance. For air temperature
measurements, check the radiation shield monthly to make sure it is free from
dust and debris. To clean the shield, first remove the sensor. Dismount the
shield. Brush all loose dirt off. If more effort is needed, use warm, soapy water
and a soft cloth or brush to thoroughly clean the shield. Allow the shield to dry
before remounting.
Periodically check cabling for signs of damage and possible moisture intrusion.
are correct for the
Therm108()
). Check the cable
(p. 5)
(p. 9)
. Ensure the clear wire is connected to
(p. 9)
, and

Advertisement

Table of Contents
loading

Table of Contents