Ossila G2001 User Manual page 17

Optical spectrometer
Table of Contents

Advertisement

Spectrometer
9.2.2 Quick Access Buttons and Acquisition Settings
Buttons to access common features. Hover the mouse cursor over the button to display a tooltip.
The buttons have the following functions:
Open previously saved spectra.
Save checked spectra.
Open device connection menu.
Continuous Data Acquisition.
Single Shot Data Acquisition.
Stop Data Acquisition.
Pin new spectra. This will keep all newly recorded spectra in the workspace.
Take a background spectrum.
Take a reference spectrum. Note that a background spectrum must be taken first.
Commonly adjusted acquisition settings are also available in this area.
(I) Integration Time
Sets the integration time of the acquisition. The allowed range is from 50 µs to 10 s (0.05 ms to
10000 ms).
(II) Delay
Introduces a delay between successive measurements. A delay of 0 seconds will result in the
spectrometer taking acquisitions as quickly as possible. The maximum allowed delay is 99.99
seconds. This value is ignored in 'external trigger' and 'rolling integration' modes.
(III) Accumulation
Sets the number of successive spectra to sum together. The maximum allowed value is 20.
(IV) Averaging
Sets the number of successive spectra that are used to calculate a mean value. The maximum
allowed value is 99.
www.ossila.com
17
User Manual
Ossila Limited © 2023