980 HDMI Protocol Analyzer Application Guide – Source Compliance Test Automation
- bit values for 6 and 7 pertain to MHL only.
-p is a parameter that pertains to HDCP and is only used for the 980 GUI Manager and not used for Automated
Test Systems.
-q is a parameter that when set to 0 indicates that only a quick test is performed on tests 7-16, 7-17 & 7-18 and no
detailed analysis of any failures is conducted. A value of 1 indicates that detailed analysis will be performed.
The default is 0.
-r
is a parameter that allows you to specify the version of the HDMI specification. Enter 1 for 1.4a and 2 for 1.3b1.
-s is a parameter that specifies the sampling rate of the incoming audio in kHz. This could be for example 48,
44.1, 96, etc.
is a parameter that provides the CTS Test ID (s) to be run. If no –t argument is provided the script will run all
-t
HDMI source tests. In the example above, only the 7-17 test will be run.
-v is a parameter provides the CEA Video Identification Code (VIC) for the format resolution being tested. If using
a non CEA format, use 0 as the value of this parameter.