Limit Alerts-Latched/Unlatched; Alert Limit Settings-Retained/Defaults - Novametrix Medical Systems CO2SMO User Manual

Etco2/spo2 monitor
Table of Contents

Advertisement

Section 8
3.
Press
The current setting flashes.
4.
Press
YES
. The user cannot use Audio Off.
NO
Once the decision to allow or disallow the user to use Audio Off is made, that choice remains in effect,
even if the monitor is turned off and on, until changed by the user.
5.
ALERT OPTIONS 1
Limit Alerts—Latched/Unlatched
Alerts caused by parameters violating alert limit settings are normally "Latched". Once a latched alert is
active for 10 seconds, even if the parameter then returns within its limits, the
indicator continue to flash until the user presses the
the alert.
CO
2
indicators as soon as the alerting parameter returns within its limits. The user does not have to press the
To select Latched or Unlatched alerts:
1.
Press and hold the
2.
Repeatedly press the
3.
Press
The current setting flashes.
4.
Press
YES
provides unlatched alerts that reset automatically without user intervention.
NO
Once the choice of Latched or Unlatched alerts is made, that choice will remain in effect, even if the
monitor is turned off and on, until changed by the user.
5.
ALERT OPTIONS 1
Alert Limit Settings—Retained/Defaults
When
monitor was last turned off. However, the monitor can be configured to use its default alert limit values
at each power up instead.
64
SMO User's Manual
CO
2
and
MUTE
ALLOW AUDIO OFF
or
as desired.
YES
NO
. The user can use Audio Off to permanently silence audible alerts.
reappears. Press
also supports "Unlatched" alerts that automatically stop the flashing of the
SMO
key when unlatched alerts are in use.
MENU
key for 3-seconds.
key until
NEXT
and
LATCH
ALERTS LATCHED
YES
or
NO
as desired.
provides latched alerts that require the user to press the
reappears
.
is turned on, it restores the (Retained) alert limit settings that were in effect when the
CO
SMO
2
appears.
AUDIO OFF DISABLED
NOTE
RUN
to return to the Main Menu.
CO2 SETUP OPTIONS
appears.
ALERT OPTIONS 1
appears.
NOTE
Press
RUN
to return to the Main Menu.
Limit Alerts—Latched/Unlatched
is displayed instead.
or
indicator and the
key. This indicates which parameter caused
is displayed.
key to clear them.
or
and
Rev. 04