Auto Alert Limits
Once the limit values are set, the user should periodically confirm patient status by alternate means
and not rely solely on alerts generated when a limit is violated.
3.
SET ALERT LIMITS
ETCO
The ETCO
recorded before
and low limits. Those ranges are:
For example, if the average of the three most recent ETCO
was 32 mmHg, the upper alert limit is set to 40 (32 + 25% = 32 × 1.25 = 40) and the lower alert limit is
set to 24 (32 - 25% = 32 × 0.75 = 24). If the average ETCO
be set to 55 (45 + 10 = 55) and the lower limit would be set to 35 (45 - 10 = 35).

Respiratory Rate Auto Alert Limits

The respiratory rate auto alert limits are derived from the average of the three most recent respiratory rate
values recorded before
high and low limits. Those ranges are:
For example, if the average of the three most recent respiration rate values recorded before
pushed was 28 br/min, the upper alert limit is set to 38 (28 + 10 = 38) and the lower alert limit is set to
21 (28 - 7 = 21). If the average respiratory rate is 12 br/min, the upper limit would be set to 19 (12 + 7 =
19) and the lower limit would be set to 6 (12 - 50% = 12 × 0.50 = 6).
SpO
2
The SpO
(maximum setting =100). The low auto alert limit is set to 5 less than the SpO
AUTO
Rev. 04
reappears. Press
Auto Alert Limits
2
auto alert limits are derived from the average of the three most recent ETCO
2
was selected. The average then falls into one of two ranges that define the high
AUTO
ETCO
2
A
Average
1 - 40 mmHg
> 40 mmHg
A.If ETCO
is displayePa or percent (not mmHg), the limits are based on
2
the equivalent of mmHg.
AUTO
was selected. The average then falls into one of three ranges that define the
Respiration
Rate Average
1 - 15
16 - 40
> 40
Auto Alert Limits
high auto alert limit is set to 5 more than the SpO
2
was pressed. (minimum setting = 50).
WARNING
to return to the Main Menu.
RUN
High
Low
Limit
Limit
+ 25% of value
- 25% of value
+ 10 mmHg
- 10 mmHg
values recorded before
2
value was 45 mmHg, the upper limit would
2
High
Low
Limit
Limit
+ 7 br/min
- 50% of value
+ 10 br/min
- 7 br/min
+ 15 br/min
- 10 br/min
value displayed when the
2
Alerts
values
2
was pushed
AUTO
AUTO
was
AUTO
was pressed
value displayed when
2
SMO User's Manual
CO
2
61

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CO2SMO and is the answer not in the manual?

Subscribe to Our Youtube Channel

This manual is also suitable for:

7100

Table of Contents