Service Documentation; Overview, Setting Regulation, Tolerances - KERN KFP 1500V40M Installation Instructions Manual

Weighing bridge
Table of Contents

Advertisement

9 Service documentation

This chapter is only intended for a balance specialist!
The weighing bridges are carried out in DMS sensor technology, at
every corner a DMS weighing cell is installed.
The analogue-digital transformation occurs in the display unit. Also all
the balance and country-specific data are stored there.

9.1 Overview, setting regulation, tolerances

13
Testing and setting regulations:
Capacity
Readability
Min
Max
1/3 corner load
Tolerance
Verification data and tolerances as per OIML
1500 kg
1
[g]
0,75
0,5
0,25
0
-0,25
-0,5
-0,75
-1
0
500
1500 kg
500 g
10 kg
1500 kg
500 kg
500 g
[kg]
1000
1500
KFP V40-IA-e-1412

Advertisement

Table of Contents
loading

This manual is also suitable for:

Kfp v40Kfp 1500v40smKfp 3000v40m

Table of Contents