Changes to the weighing system's design are not permitted. This may lead to incorrect weighing results, safety-related faults and destruction of the balance. The weighing system unit may only be operated in accordance with the described default settings. Other areas of use must be released by KERN in writing. KFP V40-IA-e-1211...
In KERN's accredited DKD calibration laboratory test weights and weighing systems may be calibrated (return to the national standard) fast and at moderate cost.
4.1 Pay attention to the instructions in the Operation Manual Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances. 4.2 Personnel training The appliance may only be operated and maintained by trained personnel.
• Keep IP protection of the device. • Major display deviations (incorrect weighing results) may be experienced should electromagnetic fields (e.g. due to mobile phones or radio equipment), static electricity accumulations or instable power supply occur. Change location or remove source of interference. KFP V40-IA-e-1211...
Ensure that the contents of package is complete. According to the version, for assembling the weighing bridge the following accessories (optional) must be available: (A) 2 access ramps (B ) 1 access ramp and 1 foot plate set (C ) 2 foot plate sets KFP V40-IA-e-1211...
In case of verified weighing systems the weighing bridge must be firmly fixed on the floor. This is essential for the reproducibility of the measuring results. Optionally using two access ramps or two pairs of foot plates or a combination of both. KFP V40-IA-e-1211...
• Avoid falling load, shock loads and impacts from the side! • For weighing operation, all objects must be placed centrically of the weighing bridge and may not hang over the edges or the ramps. • Check adjustment at regular intervals. KFP V40-IA-e-1211...
• Place the load on the scales using a lifting truck, a crane or a forklift truck. Ensure that the load is not swinging when it is placed onto the scales. • Lift the load first vertically at least 10 cm above the scales before it is removed or newly placed. KFP V40-IA-e-1211...
Do not use water jet or high-pressure cleaner. 8.3 Service, maintenance The appliance may only be opened by trained service technicians who are authorized by KERN. Ensure that the weighing system is regularly calibrated, see chap. 3.5 Testing instruments control.
• Weighing bridge on an uneven surface. • Electromagnetic fields / static charging (choose different location/switch off interfering device if possible) Should other error messages occur, switch balance off and then on again. If the error message remains inform manufacturer. KFP V40-IA-e-1211...
Testing and setting regulations: Capacity 1500 kg Readability 500 g 10 kg 1500 kg 1/3 corner load 500 kg Tolerance 500 g Verification data and tolerances as per OIML 1500 kg 0,75 0,25 [kg] -0,25 -0,5 -0,75 1000 1500 KFP V40-IA-e-1211...
• Open connection box Adjustment rule: The corner (weighing cell) with the biggest negative deviation must be set to zero. Do not re-adjust this corner even after several adjustment sequences. Adjustment on the analogue print KFP V40-IA-e-1211...
Page 15
Adjustment of weighing cell J3 takes place at the potentiometer VR2. Adjustment of weighing cell J4 takes place at the potentiometer VR3. Adjustment of weighing cell J5 takes place at the potentiometer VR4. Increase the value turning to the right, reduce the value turning to the left. KFP V40-IA-e-1211...