NEC MD205MG-1 Installation & Maintenance Manual page 14

20" grayscale 5-megapixel diagnostic monitor
Hide thumbs Also See for MD205MG-1:
Table of Contents

Advertisement

Available languages

Available languages

Quick Start
8. Hold all cables firmly and place the cable cover onto the stand (Figure E.1). To remove the cable cover, lift the cover off
as shown in Figure F.1.
9. Before operating the display, turn the Vacation Switch, located on the back of the monitor (Figure G.1) to the ON
position. After the Vacation Switch is on, then the display can be turned on/off using the power button that is also
located on the back of the display.
NOTE:
DO NOT switch on/off repeatedly.
10. The display is set to DVI "Dual Link" mode as a factory default. Power up the PC, so that the display receives a video
signal. If operation in "Single Link" mode is required, use the rear buttons on the display to change the DVI Link mode.
Press "+", "-", "ENTER", "+", "-" and select "Single Link". After changing the DVI Link mode, please reboot the PC
again.
11. For further adjustments, please refer to the OSD section for a full description of OSD controls.
NOTE:
If you have any problems, refer to the Troubleshooting section.
External equipment intended for connection to signal input, signal output, or other connectors, must comply
NOTE:
with the relevant IEC standard.
Figure E.1
01_MD205MG-1_INSTALL_EN.P65
–continued
4
Figure F.1
English.4
7/5/08, 10:09 AM
Figure G.1
Power button
Vacation
Switch