Dicom Measurement; Option - NEC MD301C4 Installation & Maintenance Manual

30" color 4-megapixel widescreen medical diagnostic monitor
Hide thumbs Also See for MD301C4:
Table of Contents

Advertisement

DICOM Measurement

1. Use "LEFT" or "RIGHT" to select DICOM MEASURE in the MODE selection (Figure D.1).
2. Select the PICTURE MODE that you want to measure and press SELECT.
NOTE:
This function is effective only if GAMMA (Advanced menu Tag1) is DICOM or PROGRAMMABLE.
3. The procedure will ask for the USB color sensor to be placed on the center of the display panel (Figure S.2). Tilt the display
panel approximately 5° backward and place the USB color sensor in the center of the display panel (Figure S.1).
NOTE:
Place the USB color sensor fl at against the display to avoid external light contamination. Do not press the USB color
sensor against the display panel. Press "SELECT".
4. After the DICOM MEASURED RESULT message appears (Figure D.2). You can set DATE and TIME using control keys,
and press SELECT
5. To end the calibration mode, press "EXIT".

OPTION

Select the STAND-ALONE CALIBRATION setting.
1. Use "LEFT" or "RIGHT" to select OPTION in the MODE selection.
2. Press the UP/DOWN button to select menu. Press the LEFT/RIGHT button to change setting.
Warning Message: A warning message will appear in STAND-ALONE calibration.
Sensor Correction: Set whether using USB Color sensor compensation (See page 30). Set "ON" normally. If the STAND-
ALONE calibration results and calibration results with other software do not match, please try to select "OFF".
Correction Table: Show the USB color sensor compensation table for current connected USB color sensor. Press SELECT
to adjust the compensation again and overwrite table.
NOTE:
Creating a compensation table is only possible within the fi rst 1000 hours (HOURS RUNNING page 22) of use.
Figure D.1
English-34
Figure D.2