Electrical connection
6.5

Connecting the sensors

Only use terminated genuine cables.
Fig. 38: Example of Memosens data cable CYK10
Connecting the ferrules of the sensor cable to the E basic module
The outer shield of the cable is grounded by means of the metal gland under the E basic module.
Fig. 39: Terminal strip
6.6
Optional sensor inputs, current outputs and relays
WARNING
!
Module not covered
No shock protection. Danger of electric shock!
If you are modifying or extending your hardware, always fill the slots from left to right. Do
not leave any gaps.
If not all the slots are occupied, always insert a dummy or end cover into the slot to the right
of the last module. This ensures the unit is shock-protected.
Always ensure shock protection is guaranteed particularly in the case of relay modules (2R,
4R, AOR).
38
a0023172
Liquiline System CA80CR
a0003350
Endress+Hauser