Instruction Manual No. 6410-00 Pen Type pH meter Model SK-610PH-II...
Page 2
Introduction Thank you for purchasing the Pen Type pH meter Model SK-610PH-II. This pH meter is designed to measure hydrogen ion concentration (pH) and temperature (liquid temperature). Do not use this instrument for any other purposes. Read this instruction manual thoroughly before using the SK-610PH-IIand keep it in a safe place for your future references.
liquids such as organic solvents, oils, adhesives, strong acids (pH 0 to pH 2), strong alkalis (pH 12 to pH 14) and surface-active agents. The pH reading may become unstable when measuring pure water. If the pH standard solution contacts the skin (hand), immediately wash the contaminated skin with running water.
Page 4
Cautions before use Check that the instrument has not been damaged during transport. If damage is found, contact the store where you purchased the instrument or our service network. When the pH of food or drink is to be measured, take a sample (sufficient quantity so that the sensor tip can be immersed at least 3 cm in the sample) in a separate container for measurement.
Page 5
Components names and functions (1) LCD : Displays measurement readings or status of the unit. key (power key) : Press this key to turn on the unit. To turn off the unit, press the key again. (3) CAL key : Enters the calibration mode. (4) HLD key : Holds the value being displayed.
Page 6
Installing and replacing battery Upper part of the (1) Take the unit in your hands as shown on the illustration on the right. body (2) Insert the tip of a Phillips screwdriver under the fixture part and release Fixture from the upper part of the body. (3) Push up the upper part of the body with your thumb to open the battery compartment.
Page 7
IIor our service network ⚫ How to calibrate the SK-610PH-II pH meter To ensure accuracy, perform calibration at the same temperature as the material to be measured. N.B. For the relation between the pH value of standard solution and the temperature, refer to the section...
Page 8
"Relation between pH value and temperature". (1) Press the key to turn on. (2) Remove the sensor cap and glass electrode cover. How to remove the glass electrode cover With the back of the unit facing you, place your fingers on the cover from sideway and turn it counterclockwise.
Page 9
Cautions The “CAL” indication does not appear if the glass electrode is dry or there is a problem with the standard solutions. Refer to “Troubleshooting” for the solution to the problem The table below shows the pH value range to be set for calibration. Calibration point pH value range pH 4...
Page 10
Do not lick or swallow the potassium chloride crystals. If the eyes or skin becomes contaminated with the crystals, immediately consult a medical attention. Caution: Spare probe model 610S-II is the exclusive probe for SK-610PH-II. Spare probe model 610S cannot be used on this.
Page 11
Troubleshooting (error codes) Problem Possible cause Action No power Batteries have worn out. Replace with new batteries (Refer to “Installing and replacing batteries”) The power was turned off The auto power-off function is on. Disable the auto power-off on the main unit. (Refer to “Before measuring”) automatically.
Specifications Cat. No. 6410-00 Pen type pH meter Model SK-610PH-II Measuring factors Hydrogen ion concentration (pH), Temperature (liquid temp.) Temperature Measuring Range 2.00 to12.00pH 0.0 to 50.0ºC (Unit and Sensor) Resolution 0.01pH 0.1ºC Accuracy ± 0.6ºC (at 20 to 30ºC) ±...