SATO KEIRYOKI SK-50TCH Instruction Manual

SATO KEIRYOKI SK-50TCH Instruction Manual

Co2 monitor

Advertisement

Quick Links

INSTRUCTION MANUAL
CO
Monitor
SK-50TCH
2
Model
No. 1737-00
SATO KEIRYOKI MFG. CO., LTD.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SK-50TCH and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for SATO KEIRYOKI SK-50TCH

  • Page 1 INSTRUCTION MANUAL Monitor SK-50TCH Model No. 1737-00 SATO KEIRYOKI MFG. CO., LTD.
  • Page 2 DO NOT USE THE INSTRUMENT IN THE AIR OTHER THAN AMBIENT ATMOSPHERE.  SK-50TCH is a precision instrument. Do not drop or give shocks to it. Do not disassemble or alter the instrument to avoid any malfunctions and damages. ...
  • Page 3: Names Of Components

    ppm higher in urbanized areas. Each breath exhaled from a person contains around 40000 ppm of thus the level in a closed room increases in general. If the level in the whole air exceeds approximately 30000 ppm, it can be harmful to life. The risk of occurrence of a cluster of infection disease arises in closed, poorly ventilated spaces.
  • Page 4: Installation

    Installation Do not install SK-50CTH in the following locations to avoid incorrect measurement in an indoor environment:  near air conditioners, heaters, vent holes, windows, or doors  where is exposed to direct sunlight, wind, or an exhaled breath Note: Keep 50 cm (2 ft) away from a nearby person at least. Flip up the stand in the rear side and put the instrument on a flat surface for table use.
  • Page 5 Settings Flowchart ⚫ Measurement mode Touch SET key for 3 seconds Setting mode Alert setting (ALr): Changes the threshold of CO alert with a red light Tap SET key Buzzer setting (bUr): Turns on or off the buzzer Tap SET key LED light brilliance setting (bL): Modifies how bright the LED light is Tap SET key * Letters in parentheses above are indicated on the humidity segment.
  • Page 6: Troubleshooting

    Recalibration of CO Sensor Accuracy of the CO sensor gets gradually worsened due to the contamination of dust and aged deterioration. For a stable long-term measurement, it is advised to recalibrate it once in six months. 1. Be sure to follow the conditions below: ...
  • Page 7: Error Displays

    Error Displays Refer to the following if error codes are displayed on any segment of the panel. If these solutions do not work, contact a dealer or us for further assistance. Segment Code Description Solution Measured value is under Use within the measuring range. the display range.
  • Page 8: Warranty Policy

    Authorization must be obtained from us before returning items for any reason. When applying for authorization, please include data regarding the reasons the items are to be returned. SATO KEIRYOKI MFG. CO., LTD. 3-4, Kanda-kajicho, Chiyoda-ku, Tokyo 101-0045 Japan URL: http://www.sksato.co.jp/english...

This manual is also suitable for:

1737-00