CLEANING INSTRUCTIONS

Cleaning the Door & Chamber
• Clean the door with a clean, lint-free cloth moistened
with isopropyl alcohol. Wipe down the surface.
• The cartridge area may be cleaned with a cotton swab
moistened with isopropyl alcohol.
Clean the Touch Screen display with a clean, lint-free cloth moistened with water
• Do not use any liquid other than water when cleaning the Touch Screen!
• To avoid damage avoid excess water on the Touch Screen!
• Never use isopropyl alcohol on the Touch Screen!
Cleaning the Exterior
• The surface of the MyCare Insite Analyser should be cleaned with a clean, lint-free cloth only moistened
with isopropyl alcohol. Wipe down the exposed surfaces.
• Follow site protocols for cleaning intervals.

TECHNICAL SPECIFICATIONS

MyCare Insite Analyser
Photometer resolution
Reproducibility
Linearity
Temperature control
Fuse
Dimensions
Weight
Communications interface RS232, USB, Bluetooth
Operating conditions
Transport/Storage
Power usage
Power Supply
Manufacturer
Type
Mains adapter
Input
Output
Power usage
SHUTDOWN AND WASTE MANAGEMENT
Before shutting down the MyCare Insite for repair or disposal, ensure that there is no Cuvette left in the Analyser
Chamber. To protect the Analyser in shipping the original packaging must be used with the transport lock installed
to prevent damage in transport.
For proper waste management according to the Directive 2012/19/EU please contact your Official Saladax
Distributor. Used cartridges need to be disposed of with laboratory waste according to the corresponding
regulations.
PI MCI-EUR-ML-PKG Rev 01
0.0001 ABS
< 1.5% CV at 1 OD
0.1000 – 3.0000 OD better than ± 1.5% and ± 0.01 OD
Electrical temperature control of the photometer unit to 37°C ± 2°C
2.5 A, self-healing
160 x 135 x 145 mm (H x W x D)
2.2 kg (unpacked)
20 - 28°C; relative humidity: 10 – 85%
A dry, clean, level surface. Avoid direct sunlight.
0 - 50°C; relative humidity: 5 – 85%
12V DC, 2A
Globtek
GTM21097-5012
A separate AC to DC mains adapter with double insulation
90 - 264V AC, 47 - 63 Hz
12V DC, 4.17A
Max. 30 VA