Step
LCD display
1
C-00
2
C-01
3
C-02
4
C-03
5
C-04
6
C-05
7
C-06
8
C-07
9
C-08
10
C-09
11
C-10
12
C-11
13
C-12
14
C-13
15
C-14
16
C-15
17
C-16
18
C-17
19
C-18
20
C-19
21
C-20
22
C-21
23
C-22
24
C-23
25
C-24
26
C-25
27
C-26
28
C-27
29
C-28
30
C-29
31
C-30
32
C-31
33
C-32
34
C-33
35
C-34
36
Save
Table 3. AAC Adjust Procedure
Calibrator output
for all models
0 A, 0 Hz
30 μA, 60 Hz
100 μA, 60 Hz
300 μA, 60 Hz
300 μA, 60 Hz
0 A, 0 Hz
300 μA, 60 Hz
1 mA, 60 Hz
3 mA, 60 Hz
3 mA, 60 Hz
0 A, 0 Hz
3 mA, 60 Hz
10 mA, 60 Hz
30 mA, 60 Hz
30 mA, 60 Hz
0 A, 0 Hz
30 mA, 60 Hz
100 mA, 60 Hz
300 mA, 60 Hz
300 mA, 60 Hz
0 A, 0 Hz
6 mA, 60 Hz
20 mA, 60 Hz
60 mA, 60 Hz
60 mA, 60 Hz
0 A, 0 Hz
60 mA, 60 Hz
200 mA, 60 Hz
600 mA, 60 Hz
600 mA, 60 Hz
0 A, 0 Hz
600 mA, 60 Hz
900 mA, 60 Hz
1.2 A, 60Hz
1.2 A, 60Hz
Push to save
AC Leakage Current Clamp
Equipment
Only use 552xA
Only use 552xA
Only use 552xA
Only use 552xA
Only use 552xA
Only use 552xA
Only use 552xA
Only use 552xA
Only use 552xA
Only use 552xA
Only use 552xA
Only use 552xA
Only use 552xA
Only use 552xA
Only use 552xA
Only use 552xA
Only use 552xA
Only use 552xA
Only use 552xA
Only use 552xA
Use 552xA and 50 turns coil
Use 552xA and 50 turns coil
Use 552xA and 50 turns coil
Use 552xA and 50 turns coil
Use 552xA and 50 turns coil
Use 552xA and 50 turns coil
Use 552xA and 50 turns coil
Use 552xA and 50 turns coil
Use 552xA and 50 turns coil
Use 552xA and 50 turns coil
Use 552xA and 50 turns coil
Use 552xA and 50 turns coil
Use 552xA and 50 turns coil
Use 552xA and 50 turns coil
Use 552xA and 50 turns coil
STBY
Calibration
13