LED Functions
Normal Operation
STATUS
Functioning - System Not Flowing
Functioning - System Flowing
Meter Malfunction
No Power / Meter Failure
Compensation via Interface Software Kit or Gold PCA Button
STATUS
Compensate button pushed/clicked
Compensation Succeeded
Compensation Failed
Note: Revolutions of meter are dependent on expected fluid flow rate. Per Table A, the maximum compensation flow
rate will take approximately 10 seconds and the minimum compensation flow rate will take approximately 90 seconds
to complete 5 revolutions at the stated flow rates.
If compensation fails, restart the compensation process. When using the Interface Software Kit, follow instructions
provided to correct errors prior to attempting compensation again. If problems persist, contact Max Machinery at
707-433-2662 or at
www.maxmachinery.com.
P-Series User Manual Rev2022
LED
Blue or Green Steady
Blue / Green alternating
colors
Red Steady
No Light
LED
Blue Steady
Green Steady
Red Steady
12
ACTION
Meter is functioning, but no flow is detected in
the system.
Meter is functioning, flow is occuring in the
system. Note: At high flow rates the alternating
Blue/Green flashes may appear as a single
steady color.
Refurbish Meter.
- Verify power supply to meter.
- Verify wiring follows diagrams on page 4.
- If power and wiring are verified, refurbish
meter.
ACTION
- Steady light for 4-5 revolutions of meter.
- Steady light for 5 seconds.
- Software will show "Success" on screen.
- Steady light for 5 seconds.
- Software will show "Flow too low/ too high/
unstable" on screen.