LabSen electrochemical sensors are premium pH electrode with manufacturing technology
and key components imported from Switzerland. LabSen833 pH electrode adopts special HF
glass membrane, suitable for strong acid solution, or solution containing HF (≥3pH)
This probe has following features:
⚫ Impact-resist membrane (see the right picture), there is no danger of electrode
breakage during normal use.
⚫ Extremely long service life, able to measure solution containing HF for about
1000 times. (test condition: 25°C, 1% HF concentration, test period 1
minute )
⚫ Blue gel inner solution, does not flow and will not cause bubble.
⚫ Silver ion trap reference prevents the contamination of the junction when
testing sample contains sulfides and proteins.
Technical Data
1
Measuring Range
Temperature Range
Shaft Material
Membrane Typ
Membrane Resistance
Reference
Junction
Usage and Maintenance
2
2.1 When measuring, please unscrew the bottle cap, pull out the electrode and rinse it with
deionized water. After using, please put the electrode back into the bottle and screw tight the
cap.
2.2 Prior to measurement, remove the rubber plug to maintain pressure of the reference solution,
LabSen 833 HF pH Electrode
User Manual
(0 ~ 12) pH
(0 ~ 100) °C
Lead-free Glass
HF
<400MΩ
Silver Ion Trap
Ceramic
Before pulling out or putting back the electrode,
make sure to fully loosen the blue cap on the
storage sleeve so that the electrode can move in
and out smoothly.
Electrolyte
Soaking Solution
Temperatursensor
Electrode Dimension
Connector
Cable
3M KCl
3M KCl
NTC 30KΩ
(Ø12×120) mm
BNC/RCA
Ø5×1m
Need help?
Do you have a question about the LabSen 833 and is the answer not in the manual?
Questions and answers