Atmos Technical Specifications; Technical Specifications - UniPix ATMOS Operation Manual

Pipette accuracy tester
Hide thumbs Also See for ATMOS:
Table of Contents

Advertisement

ATMOS Technical Specifications

Maximum accuracy error formula:
P = nominal pipette volume
P=1000
P=200
P=100
P=20
P=10

Technical Specifications

Dimensions
Weight
Power supply
Battery
Operational temp.
Altitude
± ���� = ( 4% ∗ �������� ) + 0.1
1000 µl
200 µl
100 µl
20 µl
10µl
Figure: Atmos accuracy in function of tested volume displacement
12.5 x 8.5 x 3 cm
420 g
5V DC, 2A
2000 mAh, Lithium-Polymer
from 10 °C to 40°C
from sea level up to 3500m
ATMOS Operation Manual
[ µ���� ]
Absolute accuracy
(= trueness)
±4 µl
±0.1 µl
±0.5 µl
±0.18 µl
±0.14 µl
Repeatability
(= precision)
±1.5 µl
±0.3 µl
±0.15µl
±0.05µl
±0.06µl
page 22/24

Advertisement

Table of Contents
loading

Table of Contents