Results Screen

1-channel version
Measurement
number
List of previous
measurements
(up to 10 values)
Skip: This will remove
the current
measurement (data will
not be stored)
The result screen shows the following data:
Nb: Measurement number. Up to 10 measurements can be shown.
RESULT: Measured volume displacement of each measurement.
Leak rate: Expressed in volume loss per second. A leak above 0.5%/sec of test volume indicates a leaky pipette, or
leaky tip connection (or a bad insertion of the tip into ATMOS port).
The following actions are possible:
SKIP: This will remove the current measurement. It will not be stored in the memory.
NEXT: This will start the next measurement. The current measurement will be stored in the memory.
END: This will stop the series and display statistics (possible only after 3 measurements). Usually, 3 measurements
are enough to get a good estimate of pipette performance.
ATMOS Operation Manual
Leak rate (in volume
loss per second)
Measured volume
displacement
End: Stop the series
(current data will be
stored. Average value
and STDEV will be
displayed)
Next: Start the next
measurement (current
data will be stored)
page 10/24