Serial Dilution Function - UniPix ATMOS Operation Manual

Pipette accuracy tester
Hide thumbs Also See for ATMOS:
Table of Contents

Advertisement

Serial Dilution Function

This function is designed to simplify the preparation of a serial dilution.
Click on the button "Serial Dilution" on the main screen to start a calculation.
Volume of
compound
Factors of dilution
Volume of solvent
series
Select other default
values for factors
(by 10, by 2)
Volume of final
volume
Reset all values
Go back to main
screen
The serial dilution calculator helps you calculate up to 6 concentrations.
The left, pink column shows the dilution factors. On the right, you see the volumes of compound + solvent.
The default final volume is 1500µL. To change this value, click in the number ("1500"). A number keyboard
opens and you can insert your custom volume.
Please note: the button << is "Enter".
The default calculation is a dilution by 3. By clicking on the pink button on the bottom of the display, you can
select other default factors (by 10 or by 2). If you want to use customized values, click on the pink numbers in
the left column. A number keyboard will show up and you can overwrite the default values.
The button "AC" will reset all values.
You can change the measurement units by clicking on the unit (e.g. click on "mL" to change to "µL").
ATMOS Operation Manual
page 14/24