Ai Sample Clock Timebase Signal - National Instruments PCIe-6323 User Manual

Hide thumbs Also See for PCIe-6323:
Table of Contents

Advertisement

Chapter 4
Analog Input
When using an externally generated AI Sample Clock, you must ensure the clock signal is
consistent with respect to the timing requirements of AI Convert Clock. Failure to do so may
result in a scan overrun and will cause an error. Refer to the
more information about the timing requirements between AI Convert Clock and AI Sample
Clock.
Figure 4-16 shows the relationship of AI Sample Clock to AI Start Trigger.

AI Sample Clock Timebase Signal

You can route any of the following signals to be the AI Sample Clock Timebase
(ai/SampleClockTimebase) signal:
100 MHz Timebase (default)
20 MHz Timebase
100 kHz Timebase
PXI_CLK10
RTSI <0..7>
PFI <0..15>
PXI_STAR
PXIe_DSTAR<A,B>
Analog Comparison Event (an analog trigger)
AI Sample Clock Timebase is not available as an output on the I/O connector. AI Sample Clock
Timebase is divided down to provide one of the possible sources for AI Sample Clock. You can
configure the polarity selection for AI Sample Clock Timebase as either rising or falling edge,
except on 100 MHz Timebase or 20 MHz Timebase.
4-26 | ni.com
Figure 4-16. AI Sample Clock and AI Start Trigger
AI Sample Clock Timebase
AI Start Trigger
AI Sample Clock
AI Convert Clock Signal
Delay
From
Start
Trigger
section for