RANGER
7000 SCALES
5.5

Read Out

Enter this sub-menu to customize Scale display functionality.
Scale Read Out sub-menu

Reset

Auto Zero
Tracking
Factory default settings are shown below in bold.

Stability

Zero Range

Brightness
Auto Dim
(minutes)
5.5.1
Reset
Reset all settings to factory default settings.
Yes = Reset.
No = Do not reset and return to Read Out menu screen.
5.5.2
Stability
Set the amount the reading can vary while the stability symbol remains on.
0.5 Division = 0.5 graduations
1 Division = 1 graduation
2 Division = 2 graduations
5 Division = 5 graduations
Note: The setting is forced and locked to 1 Division when the Security Switch
is set to the locked position.
5.5.3
Zero Range
Set the percentage of scale capacity that may be zeroed.
2%
10%
Note: The setting is forced and locked to 2% when the Security Switch is
set to the locked position.
5.5.4

Filter level

Set the amount of signal filtering.
LOW
= faster stabilization time with less stability.
MEDIUM
= normal stabilization time with normal stability.
HIGH
= slower stabilization time with more stability.
Note: The setting is at the current setting when the Security Switch is set to
the locked position.
Filter Level
Auto Sleep
(minutes)
EN-45