RT‑ZM
Operating voltage window
(each pin to GND)
Maximum non-destructive input voltage ±30 V DC
Input resistance
R&S ProbeMeter, measurement error
R&S ProbeMeter, dynamic range
Offset capability
Low zero and gain errors throughout the entire temperature range for all measurement modes
Micro button
Rohde & Schwarz probe interface
2.1.2
Key features
Micro button
The micro button remotely controls important functions of the Rohde & Schwarz
oscilloscope. The function is assigned at the oscilloscope.
For details, see
Chapter 4.4, "Micro
R&S ProbeMeter
The R&S ProbeMeter measures the DC voltage of the input signal directly at the
probe tip. It provides a continuous high-precision DC voltage measurement that is
independent of the settings of the oscilloscope and runs in parallel to the time
domain measurement. If activated on the base unit, the measured value is dis-
played on the screen of the Rohde & Schwarz oscilloscope.
The R&S ProbeMeter simultaneously measures both differential and common
mode DC voltages.
For details, see
Chapter 4.5, "R&S
Probe DC correction
The probe amplifier includes an integrated data memory with individually deter-
mined DC correction parameters (e.g. gain, offset). These correction parameters
are read out and processed by the Rohde & Schwarz oscilloscope. As a result,
User Manual 1419.3128.02 ─ 05
Key characteristics and key features
±7.0 V with ±16 V offset capability, DC to 100 kHz
±5.0 V with ±16 V offset capability, > 100 kHz
Differential: 400 kΩ
Single-ended: 200 kΩ
<0.2 %
±7 V
±16 V for compensating single-ended, differential and
common mode DC offsets