NiCaVis 70x IQ NI
Cleaning agents
Accessories
ba75988e01
11/2012
5
Maintenance and cleaning
5.1
Maintenance
The UV NiCaVis 70x IQ NI sensor operates maintenance-free.
5.2
Sensor cleaning
5.2.1 Cleaning agents and accessories
To clean the sensor, use the following cleaning agents only:
Contamination
Water-soluble substances
Lime deposits
Grease deposits
* For the final cleaning prior to a
use cleaning agents with purity grade "for analysis". Finally, thoroughly
rinse the sensor with ultrapure water.
Sponge
Plastic-bristled brushes of different sizes.
Interdental brushes are suitable for the cleaning of the compressed
air channels.
Flocked cleaning cards (see accessories)
Lens tissues
Wash bottle
Note
Never use any pressure washer, steam blaster, abrasives or metal
brushes to clean the sensor. Risk of damaging the sensor.
Maintenance and cleaning
Cleaning agents
– Primary cleaning: Tapwater,
warm if possible
(30 - 50 °C)
– Precision cleaning: House-
hold washing-up liquid + ultra-
pure water.
5 % hydrochloric acid *
Mixture of isopropanol and water
(approx. 70 % isopropanol) *
Sensor check
Zero adjustment
or
, only
5 - 1