Calibration; Overview; User Calibration; Sensor Check/Zero Adjustment - wtw IQ SENSOR NET NiCaVis 705 IQ NI Operating Manual

Table of Contents

Advertisement

NiCaVis 70x IQ NI

User calibration

Recalibration after
changing the measuring
location
ba75988e01
11/2012
4.3

Calibration

4.3.1 Overview

In the delivery condition, the sensor is precalibrated and ready to mea-
sure (see section 3.4.1).
There are two types of calibration that are included into the calculation
of the measured value at different points (see section 4.1):
 User calibration (see section 4.3.2)
Zero adjustment
 Sensor check/
An overview diagram of the calibration procedures can be found on
page 6 - 9.
With a user calibration you compensate for the influence of the test
sample. The user calibration requires a successful
dest.)
Zero adjustment
and a valid
We recommend to check the current user calibration against indepen-
dent comparison measurements if you suspect a basic change of the
characteristics of the test sample. If the comparison measurements
show clear deviations, these can be compensated by a user calibration.
Each time after the application or measurement location
(and the setting,
recommend to carry out a
(and, if the
an additional
quent user calibration.
The data of the user calibration are stored in the IQ S
troller and thus assigned to the measuring location (not to the sensor).
The calibration data of the
(see section 4.3.3).
.
Measuring location
Sensor check (H2O dest.)
Sensor check (H2O dest.)
Zero adjustment
if necessary) with a subse-
Zero adjustment
Measuring / Operation
Sensor check (H2O
) was changed, we
was not successful,
N
ENSOR
ET
are stored in the sensor.
con-
4 - 3

Advertisement

Table of Contents
loading

This manual is also suitable for:

Iq sensor net nicavis 701 iq

Table of Contents