To Make Time Measurements Automatically - Agilent Technologies 54610B User's And Service Manual

Table of Contents

Advertisement

Operating Your Oscilloscope

To make time measurements automatically

To make time measurements automatically
You can measure the following time parameters with the oscilloscope:
frequency, period, duty cycle, width, rise time, and fall time. The following
exercise guides you through the Time keys by making a rise time
measurement. Figure 2-6 shows a pulse with some of the time measurement
points.
Connect a signal to the oscilloscope and obtain a stable display.
1
When the signal has a well-defined top and bottom, the rise time and fall time
measurements are made at the 10% and 90% levels. If the oscilloscope
cannot find a well-defined top or bottom, the maximum and minimum levels
are used to calculate the 10% and 90% points. These levels are shown on
page 2-19 in figures 2-8 and 2-9.
Figure 2-6
2-16