Adjusting the Oscilloscope

To adjust the high-frequency pulse response

In this procedure you adjust the high-frequency pulse response for each
channel.
Equipment Required
Critical specifications
Rise time < 175 ps
SMA (f) to BNC (m)
Connect the pulse generator to channel 1.
Press
.
Autoscale
Change the time base to 10 ns/div.
Press
, then toggle the
1
Adjust the Volt/Div until there are about 6 divisions of vertical
deflection.
3-26
Recommended
Model/Part
PSPL 1107B TD and
PSPL 1110B Driver
Agilent 1250-1787
Vernier
softkey to On.