ARM-EVAC® 250 & 200 SYSTEMS
INSTALLATION
Connecting the 45mm (1.75") and 50mm (2") Static-Safe Metal Flex Arms
1) Attach the Static-Safe Metal Flex Arm (either size) at a convenient location on the workbench
either permanently or with the bench mounting bracket (P/N 8886-0552).
Permanent Mounting: The extraction arm can be permanently mounted to a workbench or
tabletop.
A.) Drill a 50mm (2") diameter hole in the workbench.
B.) Locate the Static-Safe Metal Flex Arm flange in the hole and mark the position of the 4 holes in
the mounting flange. Remove the Static-Safe Metal Flex Arm and drill the 4 marked areas with a
2mm (3/32") drill. Refit the Static-Safe Metal Flex Arm. Secure the mounting flange to the
workbench with the 4 mounting screws (supplied).
Portable Bench Mounting: The Static-Safe Metal Flex Arm can be bench mounted with the bench
mounting bracket (P/N 8886-0552).
A.) Open the bench mounting bracket butterfly screw, position the bracket over the edge of the
work bench, then tighten the butterfly screw until the bracket is securely mounted on the work
bench.
B.) Install the Static-Safe Metal Flex Arm to the bench
mounting bracket with the 4 mounting screws (supplied).
2) Position the Static-Safe Metal Flex Arm as needed.
The Arm can be repositioned simply by grasping the
Static-Safe Metal Flex Arm above the fume scoop and
moving it to the desired location.
3) Connect the flexible hose (P/N 8886-0299) to the inlet
port of the Arm-Evac 250/200 and secure it with the hose
clamp (provided). Connect the other end of the flex hose,
with the reducer, to the inlet of the Static-Safe Metal Flex Arm.
4) If only one inlet port on the Arm-Evac is used make sure the unused inlet port has its cap in
place. 5) Repeat steps 1 through 4 to install additional Static-Safe Metal Flex Arms.

ARM-EVAC 200

As the system is used, the filter will eventually become clogged with collected particulates. If air
flow is insufficient to collect the fumes, the filter(s) should be replaced. Refer to "Filter
Replacement", page 12.
1) Turn the power switch to the On ("I") position.
2) Position the Arm-Evac accessory in a convenient and effective extraction location.
Orion Industry – 16 chemin de la Guy – 91160 Ballainvilliers – France - Tel +33 1 69 34 53 11