Connection Data - Endress+Hauser Liquiphant FTL51B Safety Instruction

Hide thumbs Also See for Liquiphant FTL51B:
Table of Contents

Advertisement

XA01810F-C

Connection data

22
Optional specification, ID Nx, Ox = NF, NG
When using the Bluetooth® module: No changes to the connection
values.
Associated intrinsically safe power supply unit with max. electrical
specifications below the characteristic values of the electronic inserts
Basic specification,
Position 3, 4
A7
A8
Only associated intrinsically safe power supply unit FML621 from
Endress+Hauser
Basic specification,
Position 3, 4
GA
Cable entry parameters
Ex ia IIC
Not relevant.
Liquiphant FTL51B, FTL63
Power supply circuit
U
= 14.6 V
i
I
= 100 mA
i
P
= 633 mW
i
L
= 0
i
C
= 3 nF
i
U
= 16 V
i
I
= 52 mA
i
P
= 170 mW
i
L
= 0
i
C
= 30 nF
i
Power supply circuit
U
= 27.6 V
i
I
= 93 mA
i
P
= 640 mW
i
L
= 3 µH
i
C
= 3 nF
i