wtw TriOxmatic 700IN Operating Manual page 36

D. o. probe
Table of Contents

Advertisement

What to do if ...
The probe has been
mechanically
damaged
Measured values
too low
36
Cause
Cause
– Membrane contaminated
– Probe has not been
calibrated for a lengthy
period of time
– Membran nicht an Platin-
Arbeitselektrode anliegend
– Shunt resistance caused by
moisture in the probe, plug
or cable
Remedy
– Send probe to WTW
Remedy
– Perform an exterior
cleaning and calibration
according to the
operating manual
– Replace the membrane
head and calibrate
according to the
operating manual
– To check, disconnect
the probe, screw off the
membrane head,
thoroughly rinse the
electrode unit with
distilled water and
completely dry it. -
Connect the probe to
the instrument and
switch to the highest
measuring range (0 -
600 %): If the display is
> 1 digit: There is a
shunt. Send the probe
to WTW for repair.