Maintenance, cleaning, disposal, spare parts
4 Put the probe in the stand.
Immerse the electrode unit into the RL-AG/Oxi
cleaning solution up to a little bit above the counter
electrode.
Note
At this point the cleaning procedure deviates from the
common maintenance procedure for WTW D. O. probes!
This procedure applies for the TriOxmatic
Allow the cleaning solution to react for 1 hour.
5 Thoroughly rinse the probe head and the electrode
unit with deionized water several times.
6 Immerse the probe head and the electrode unit into
deionized water.
â
700IN only.
27