The TrueLife Care Q8 thermometer utilizes infrared radiation to achieve instant highly-accurate results. The Care Q8 thermometer is designed to measure the temperature on a person's forehead. The thermometer has two modes of measurement: body temperature and surface temperature (the thermometer can also measure room temperature, the temperature of food, bath water, baby milk and more).
Display Layout
![]() | Sound On | °F | Degrees Fahrenheit |
Body temp | Temperature Measurement Mode | 888.8 | Measurement Value |
Object temp | Object Measurement Mode | ![]() | Memory Value |
°C | Degrees Celsius | ![]() | Low Battery Indicator |
Before Starting:
Before taking a reading, first insert the batteries. Slide the battery cover in the direction indicated by the arrow to open. Insert AAA batteries into the thermometer according to the polarity indicated. Then close the battery cover until you hear a click.
Wait 10 minutes for the device to warm up before using it for the first time. If the device has not been used for an extended period of time, it will measure room temperature for 1–2 seconds after switching on.
Taking a Reading:
Press the „Power On / Measure" button to turn on the thermometer. Point the thermometer towards the forehead at a distance of 5–15 cm, press the measure button and the temperature will be displayed. Make sure you have no obstacles such as hair, cosmetics, sweat or headgear in the measurement path.
If you cannot measure the temperature from the forehead due to excessive sweating, measure the temperature from behind the ear. Again, make sure that nothing is interfering with your measurements. For the most accurate results, make sure the room temperature is stable. Readings taken in front of a fan, air conditioner, or in a room where air flows at different temperatures may affect the result. The reading will not be accurate if the patient is using a cold compress. It is recommended to take the temperature 2-3 times in succession to ensure that the result is not affected by external factors.
After turning on the device, press the „Change Mode" button to switch between body temperature and surface temperature measurement. The selected mode will be displayed on the thermometer display. Make sure you have the mode set correctly.
Enter advanced settings by pressing the „Settings" button on the side of the thermometer. We recommend that you keep the default settings as the device is properly calibrated. If changes are made there is a risk of inaccurate results. However, if you consider it necessary, you can change the following settings:
F-1 | Setting the unit of measurement | Select whether you want your readings in Celsius or Fahrenheit. The device is preset to use degrees Celsius. |
F-2 | Mute settings | Select whether you want the device to beep when you take a reading. |
F-3 | Setting the temperature alert | Change the temperature value at which the device will give you a high temperature alert. The default value is 38°C. |
F-4 | Sound Settings | Measurement variation adjustment - If necessary, you can adjust the measurement by +/- 2.0°C in tenths of a degree. Press the ‚+' and ‚-' buttons to achieve the desired result. |
To change settings, press the "Settings" button repeatedly until SAVE is displayed. Then OFF will be displayed and the device turns off.
The low battery indicator will illuminate when the battery is running low on power.
Remove weak batteries from the thermometer. Do not leave them in the device as the thermometer may be damaged.
Insert new AA batteries into the thermometer according to the polarity indicated. Then close the battery cover until you hear a click.
When disposing of old batteries, follow the safety information instructions enclosed in the package, or hand it over to a collection point or a place designed for this purpose.
Model: | TrueLife Care Q8 |
Measurement range: | Body temperature (32°C – 42.9°C), Ambient temperature (0°C – 100°C) |
Measurement deviation: | +/- 0.1°C |
Measurement distance: | 5–15 cm |
Measurement time: | 0.5 s |
Battery: | 2× 1.5V AA |
1× TrueLife Care Q8
2× 1.5V AA
1× Manual
Reading is not as expected | The thermometer should be used in stable conditions. Measurements may show deviations when ambient conditions change frequently or when very hot and cold objects are measured within a short space of time. |
Display shows Hi | Measured temperature is higher than 42.9°C when measuring body temperature or above 100°C when measuring ambient or object temperature. |
Display shows Lo | Measured temperature is below 32.0°C when measuring body temperature or below 0°C when measuring ambient or object temperature. |
Importer:
elem6 s.r.o., Braškovská 15, 161 00 Prague, Czech Republic Made in P.R.C.
The indicated symbol (crossed-out bin) on the product or accompanying documentation means that used electrical or electronic products must not be disposed of with municipal waste. In order to dispose of the product correctly, please return it to the designated collection point where it will be accepted for free. Manufactured by: Shenzhen Brav Electronic Technologies Co., Ltd Add 4/F, Block 11, Tongfuyu Industry Park, Lezhujiao, Xixiang, Shenzhen. Use your dealer for warranty repairs. If you have any technical problems or questions, please contact your dealer for further information. The product warranty period is 24 months unless otherwise specified.