Table 11. Guidance and Manufacturer's Declaration –
Electromagnetic Immunity
Immunity
Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Conducted RF
IEC 61000-4-6
IEC 60601
Test Level
±8 kV contact
±15 kV air
±2 kV for power
supply lines
±1 kV for input/
output lines
±1 kV line to line
±2 kV line to
ground
<5% U
T
40% U
T
70% U
T
30 A/m
3 Vrms
150 kHz to
80 MHz
86
Compliance
Level
±8 kV contact
±15 kV air
Not applicable
Not applicable
Not applicable
30 A/m
3 Vrms
Electromag-
netic
Environment
Guidance
The relative
humidity should
be at least 15%.
No guidance for
battery-powered
devices.
No guidance for
battery-powered
devices.
No guidance for
battery-powered
devices.
Power frequency
magnetic fields
should be at
levels
characteristic of
a typical location
in a typical
commercial or
hospital
environment.
The device is
intended for use
in the
electromagnetic
environment in