P-2000
o w n e r ' s m a n u a l
W H E N A C C U R A C Y I S T H E P O I N T
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
®
.
TM

Advertisement

Table of Contents
loading

Summary of Contents for Delmhorst Instrument Co P-2000

  • Page 1 P-2000 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.TestEquipmentDepot.com o w n e r ’ s m a n u a l ® W H E N A C C U R A C Y I S T H E P O I N T...
  • Page 2: Table Of Contents

    TABLE OF CONTENTS 2 P-2000 Features 3 Key Functions 3 Check Calibration 4 To Set the Scale 4 To Change the Set-Point 5 Information About Your Readings 5 To Check the Accumulated Readings 6 To Reset the Meter 6 Applications...
  • Page 3: 2000 Features

    Read Key Calibration Check Key Scale P-2000 FEATURES h Designed to check the moisture levels in paper products such as cores, corrugated, kraft stock, baled scrap paper, and other paper-based materials h Resistance technology recognized around the world as the...
  • Page 4: Key Functions

    KEY FUNCTIONS Read: Press this key to read the %MC or relative moisture value of the material under test. Check: This key, when pressed with the Read key, checks the meter calibration. It also displays the number of readings in memory (up to 100), the average, and the highest stored reading.
  • Page 5: To Set The Scale

    Disconnect the battery. Press and hold the Read key for 15 sec- onds. Release the Read key. Press and hold the Check key for 15 seconds. Release the Check key. Connect a fresh battery to the lead wire in a single action, making sure to align the poles prop- erly and without interrupting contact.
  • Page 6: Information About Your Readings

    INFORMATION ABOUT YOUR READINGS Readings below the nominal range of each scale will be dis- played as a negative number. Readings above the nominal range will be displayed with a blinking number. All under- range and over-range readings should be disregarded. They are not added to the accumulated readings or used in calcu- lation of the average or highest reading.
  • Page 7: To Reset The Meter

    TO RESET METER hPress and release the calibration check key hWithin one second, press and hold the scale key . The meter will display a reset sequence as follows: “141”, “7”, “2.0”, “11.1”. The last number, “11.1” is a cali- bration check.
  • Page 8: Using The 0-100 Arbitrary Scale

    Using the 0 –100 Arbitrary Scale This scale is used to test the moisture content of hygroscopic materials for which a calibration is not available. Depending on the material, a special application external electrode, instead of the integral contact pins may be required. Increasing readings on the 0-100 reference scale indicate higher levels of moisture content.
  • Page 9: Care Of Your Meter

    A few meter readings in a limited number of specific areas of a large mass can not be projected to indicate an average moisture content of an entire bale.The readings can be very helpful in providing an indication of the overall moisture con- dition inside the bale and to detect areas of excessive mois- ture.
  • Page 10: Warranty

    90 days. If, within the war- ranty period of the P-2000, you find any defect in material or workmanship return the meter following the instructions in the Service for Your Meter section.
  • Page 11 NOTES...
  • Page 12 For over 65 years Delmhorst Instrument has been the leading manufacturer of high quality, US- made moisture meters and ther- mo-hygrometers. Today we offer a wide range of meters for appli- cations including water damage restoration, construction, floor- ing, lumber/woodworking, paper, and agriculture.