C-2000 manual_2013-Revs_FX-2000 owners manual 1/25/13 11:17 AM Page 4 TABLE OF CONTENTS 2 C-2000 Features 3 Key Functions 3 Check Calibration 4 Set the Scale 4 To Change the Set-Point 5 Information About Your Readings 5 To Check the Accumulated Readings...
C-2000 manual_2013-Revs_FX-2000 owners manual 1/25/13 11:17 AM Page 5 D E L M H O R S T C - 2 0 0 0 Set-Point Read Key Calibration Check Key Scale C-2000 FEATURES hDesigned primarily to check the moisture content (%MC) in lint cotton and seed cotton.
C-2000 manual_2013-Revs_FX-2000 owners manual 1/25/13 11:17 AM Page 6 KEY FUNCTIONS Read: Reads the % MC value. Check: Checks the meter calibration. Displays the average of up to 100 accumulated readings; displays the highest stored reading; erases the readings from memory.
C-2000 manual_2013-Revs_FX-2000 owners manual 1/25/13 11:17 AM Page 7 A hard Reset is required if, after changing the battery, the display is frozen. This is sometimes caused by the interrup- tion of contact between the battery and the battery lead wire.
C-2000 manual_2013-Revs_FX-2000 owners manual 1/25/13 11:17 AM Page 8 INFORMATION ABOUT YOUR READINGS The meter will accumulate up to 100 readings. After all 100 readings are “stored” it will not add new readings until the memory has been cleared. It will also continue to display the average of all 100 readings as a reminder that the memory is full.
C-2000 manual_2013-Revs_FX-2000 owners manual 1/25/13 11:17 AM Page 9 hTo keep the accumulated readings in memory release the cal- ibration check key before the total cycle time is complete. hTo erase readings, hold the calibration check key more than five seconds. The total, average and highest read- ings will be displayed as above, followed by a zero to indicate all readings have been erased.
Page 9
C-2000 manual_2013-Revs_FX-2000 owners manual 1/25/13 11:17 AM Page 10 Readings below 6% seed cotton will be displayed as a nega- tive number. Readings above 20% seed cotton will be dis- played as a flashing “20.0%”. Both the underrange and the overrange readings should be disregarded.
Page 10
C-2000 manual_2013-Revs_FX-2000 owners manual 1/25/13 11:17 AM Page 11 Testing Baled Cotton hAttach the 30-E/C Electrode to the connector on top of the meter. hSlide the plastic spacer over the 491 contact pins from the bottom of the pins, mount the pins in the chucks and tighten the set screws.
C-2000 manual_2013-Revs_FX-2000 owners manual 1/25/13 11:17 AM Page 12 Temperature Correction The basic calibration assumes the cotton temperature to be 70°F. For best accuracy, apply a temperature correction if the cotton temperature is outside the range of 60°F to 80°F.
C-2000 manual_2013-Revs_FX-2000 owners manual 1/25/13 11:17 AM Page 13 SERVICE FOR YOUR METER If your meter is not working properly, replace the battery with a new one and check the calibration. If this does not resolve the problem, go to www.delmhorst.com and follow the instructions under the Product Support tab.
90 days. If, within the war- ranty period of the C-2000, you find any defect in material or workmanship return the meter following the instructions in the “Service for Your Meter” section. This limited war-...
Page 16
C-2000 manual_2013-Revs_FX-2000 owners manual 1/25/13 11:17 AM Page 1 For 65 years Delmhorst Instrument has been the leading manufacturer of high quality, US- made moisture meters and thermo- hygrometers. Today we offer a wide range of meters for applica- tions including water damage...
Need help?
Do you have a question about the C-2000 and is the answer not in the manual?
Questions and answers