Model Features; Operating Features - Welch Allyn SureTemp Plus 692 Series Service Manual

Hide thumbs Also See for SureTemp Plus 692 Series:
Table of Contents

Advertisement

2.2

Model Features

Features of the SureTemp
Table 2-1: Feature differences between Models 690 and 692
Feature
Probe Cover Storage
F/C Conversion
Last Temperature Recall
Monitor Mode
Detachable Probe
Oral/Axillary/Rectal Modes on Each
Thermometer
Color-Coded Removable Probe Well
Wall-Holder
Pulse Timer
Backlight
Security (Electronic)
On-Screen Labeling for Instrument
Identification
2.3

Operating Features

2.3.1 Instrument Reset
To perform a reset:
1. Remove the batteries from the instrument to reset the internal microprocessor electronics.
Follow the battery removal instructions in "Replacing the Batteries" on page 16.
2. Remove the probe and the probe well from the instrument. Then, unplug the probe
connector by depressing the locking tab and pulling on the connector body. DO NOT pull
on the cord.
3. While watching the display, install the batteries per the instructions in "Replacing the
Batteries" on page 16 and observe the Power-On Self Test described below.
2.3.2 Power-On Self Test (POST)
The self-test will check the internal microprocessor,
instrument electronics, and display for correct operation.
The instrument performs the self-test series following all
power-on or reset cycles. During the self test sequence the
thermometer displays all LCD segments for 2 seconds.
Concurrently, the thermometer plays the power-on audible
notification and performs necessary self-testing.
SureTemp® Plus Service Manual
®
Plus Thermometer models (690 and 692) are listed below.
Model 690
Optional
Model 692
X
X
X
X
X
X
X
X
X
X
X
X
X
X
Standard
X
X
X
X
Figure 2: Model 690 and 692
Thermometer Displays
7

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Suretemp plus 690 series